Forcing a calibration line through zero needs evidence

2 min

A zero concentration does not guarantee a zero instrument signal. Background response, electronic offsets and sample preparation can introduce an intercept that is small yet important near the bottom of a range.

Explanation

A line constrained through zero has fewer adjustable parameters. That can reduce numerical variability while introducing bias when the constraint is false. A visually neat graph cannot establish that the physical constraint is valid.

Example

Consider a teaching model with a response of two units before any analyte contribution. A zero-intercept fit must distribute that offset into the estimated slope and the residuals.

y = 2 + 5x

In practice

Compare blank evidence, residual patterns and independent check results before imposing a zero intercept. Record the scientific reason for the constraint as part of the calibration method.